Abstract
This single-operator phantom study demonstrates the technical feasibility of colocating a live bronchoscope video stream with patient-specific 3D airway models on Apple Vision Pro using a native Network Device Interface pipeline, without establishing comparative or clinical benefit.
JMIR XR Spatial Comput 2026;3:e103880doi:10.2196/103880
Keywords
Introduction
Bronchoscopic biopsy of peripheral pulmonary lesions remains technically demanding because the operator must mentally translate 2D computed tomography (CT) images into the 3D bronchial route.
Mixed-reality (MR) head-mounted displays ease this load by keeping a positionable 3D reference in view during the procedure. Our group has used optical see-through MR (using HoloLens 2 by Microsoft and Holoeyes MD by Holoeyes Inc) for intraprocedural 3D holographic virtual bronchoscopy [,]. Separately, without this platform, we have shown the feasibility of Apple Vision Pro (AVP) for simulation-based bronchoscopy training [].
We developed a prototype visionOS application that ingests the bronchoscope feed via the Network Device Interface (NDI; Vizrt Group) and renders it as a plane colocated with the patient-specific 3D airway model. We report a single-operator phantom feasibility study as a proof of concept.
Methods
Ethical Considerations
Existing clinical CT images of patients at Fujita Health University Hospital were used retrospectively and anonymized to generate the models. The institutional review board of Fujita Health University approved the study (HM25-343) and waived written informed consent, with an opt-out procedure, in accordance with the Japanese Ethical Guidelines for Medical and Biological Research Involving Human Subjects. No procedures were performed on patients or volunteers.
Application
We established a native visionOS 2 application (Swift, SwiftUI, RealityKit) that receives a live video stream over NDI through a custom Swift bridge to the NDI SDK for Apple Silicon and renders it colocated with patient-specific 3D airway models. Two STL meshes per case—the airway tree (pink) and the target nodule (green)—are loaded at runtime. Using ARKit hand tracking, the video plane and each model can be independently repositioned, rescaled, or hidden. No third-party MR platform was used.
Datasets
Ten patient-derived datasets, each a different bronchopulmonary segment, were generated from thin-slice chest CT images (0.5‐1.0 mm) on a workstation (Ziostation REVORAS, Ziosoft) using the same pipeline as our previous MR virtual bronchoscopy work [,].
Bench Setup and Protocol
An aScope 5 Broncho bronchoscope (4.2/2.2; Ambu A/S) was advanced in a Koken LM-099 airway phantom (Koken). Its image was passed by HDMI (High-Definition Multimedia Interface) cord from the Ambu monitor to a dedicated computer running the Ambu Broncho Simulator (version 2; Ambu A/S), a platform that labels bronchial segments. The simulator output was captured using a USB HDMI capture card (CM489; UGREEN) on a MacBook Air (M2; Apple Inc) and sent to AVP through OBS Studio with the DistroAV plug-in via NDI over a 5-GHz personal hotspot (iPhone 16; Apple Inc). The experimental setup is shown in and the operator’s view is shown in and .

A single board-certified bronchoscopist (SO, 20 years of experience) performed 10 trials in one session in random segment order. For each trial, the corresponding dataset was loaded, and the operator advanced the bronchoscope to the target segmental orifice while viewing the NDI video plane and the 3D model on the AVP. The operator’s view is shown in .
Outcomes
Primary outcomes were per-trial time-to-target and reach success (binary). Time-to-target was read from the simulator’s built-in timer, which started when the scope passed the carina and was recorded when the target segmental label changed color. Secondary outcomes were operator-reported Simulator Sickness Questionnaire (SSQ) scores before and after the session [] and a free-text log of instability. Transport latency was measured separately (). No statistical inference was performed.
Results
All 10 target segments were reached, for a per-trial reach success of 10/10 (100%). The mean time-to-target was 37.5 (SD 5.0; range: 30‐44) seconds.
The integrated workspace remained operable throughout the session, which was completed without removing the headset. A single transient video freeze occurred during one trial (R5), from which the stream recovered immediately; its cause was subsequently identified as unbounded frame queuing in the NDI receiver and corrected. No loss of control, image corruption, or other instability occurred in the remaining trials. Latency was not measured during the trials; in bench measurements made 3 months later, on a subsequently modified build, latency of the NDI path relative to pass-through was 250 (median; IQR 232‐261) milliseconds, approximately 261 milliseconds from sensor to display ().
Total SSQ severity was 0 before and 7.5 after the session. The increase was confined to the oculomotor domain (general fatigue and eyestrain, each rated slight). No nausea or disorientation was reported, and no adverse events occurred.
Discussion
To the best of our knowledge, this is the first report to colocate a live bronchoscope video stream—ingested through a native NDI-based pipeline, without a third-party MR platform—and a patient-specific 3D airway model in a single immersive scene on AVP.
Our work shares a live-video approach with recent surgical applications of AVP [,] but applies it to peripheral bronchoscopy and colocates the endoscopic image with the patient-specific 3D airway model in one scene rather than beside separate 2D windows.
This study included no comparator, neither conventional bronchoscopy nor the 3D model without colocated video, and therefore cannot determine whether the integrated display confers any benefit over existing approaches; all findings are descriptive. The 10/10 reach success in a static phantom by a single experienced bronchoscopist is near-ceiling by design, cannot be attributed to the system, and is not evidence of efficacy. Target reach was assessed only to the segmental bronchus, the level the simulator labels. Workload was not assessed beyond the SSQ. The measured latency is not negligible; it was obtained 3 months after the trials on a modified build, at a single short range, and excluded the Ambu Broncho Simulator stage present during the trials, so it underestimates that configuration. A dedicated low-latency network and further reduction are prerequisites for clinical use. If the stream is interrupted, the video plane stalls while the locally rendered 3D model and the bronchoscope’s own monitor remain available, though this fallback was untested.
Future studies should include a controlled comparison against conventional bronchoscopy and against the 3D model without colocated video, a multioperator study with blinded timing, and an approved first-in-human case with a coregistered 3D model.
Acknowledgments
The authors want to thank Ambu K.K. (Tokyo, Japan) for the technical support in setting up the Ambu Broncho Simulator and Professor Maki Sugimoto (Innovation Lab, Teikyo University Okinaga Research Institute, Tokyo, Japan) for providing early discussions on spatial computing approaches in bronchoscopy. The authors would like to thank Enago for the English language review by professional human editors.
Generative AI Use
Generative AI was used for two purposes in this work. The prototype visionOS application was developed by the first author with AI-assisted code generation (Rork Max, Rork Inc, and Claude Code, Anthropic) in May and June 2026; all generated code was reviewed and verified on the target hardware before use. Claude (Anthropic) was used to assist with the drafting and editorial revision of the manuscript text in June and September 2026. No generative AI tool was used to generate, analyze, or interpret study data, to produce the figure, or to draft the results. The authors reviewed and edited all output, take full responsibility for the content of the manuscript, and confirm that no AI tool is listed as an author.
Funding
This work was supported by JSPS KAKENHI (Japan Society for the Promotion of Science Grants-in-Aid for Scientific Research; grant number: 25K15218).
Data Availability
The datasets generated and analyzed during the current study are not publicly available because they contain patient-derived imaging data but are available from the corresponding author on reasonable request.
Authors' Contributions
Conceptualization: SO
Methodology: SO
Software: SO
Investigation: SO
Writing – original draft: SO
Writing – review and editing: TH, YG, MF, KI, NH
Supervision: TH, YG, MF, KI, NH
All authors read and approved the final manuscript.
Conflicts of Interest
None declared.
Multimedia Appendix 1
Representative first-person recording of the operator’s view through Apple Vision Pro during use of the integrated display, showing the live bronchoscope video colocated with the patient-specific 3D airway model. This is an illustrative demonstration recorded separately from the timed trials.
MOV File, 50681 KBMultimedia Appendix 2
Latency measurement: rationale, method, conditions, results, and limitations.
DOCX File, 39 KBReferences
- Okachi S, Sugimoto M, Ina T, et al. Clinical applicability of three-dimensional holographic virtual bronchoscopy with mixed reality. Ann Am Thorac Soc. Apr 2025;22(4):609-611. [CrossRef] [Medline]
- Ina T, Okachi S, Kamenosono S, et al. Initial experience with three-dimensional holographic virtual bronchoscopy using mixed reality for peripheral pulmonary lesions: a retrospective feasibility study. Respiration. Mar 13, 2026:1-11. [CrossRef] [Medline]
- Okachi S, Ina T, Horiguchi T, Goto Y, Hashimoto N, Imaizumi K. Navigational bronchoscopy training with spatial computing: a simulation-based feasibility report using Apple Vision Pro. Surg Innov. Apr 10, 2026:15533506261441953. [CrossRef] [Medline]
- Kennedy RS, Lane NE, Berbaum KS, Lilienthal MG. Simulator Sickness Questionnaire: an enhanced method for quantifying simulator sickness. Int J Aviat Psychol. Jul 1993;3(3):203-220. [CrossRef]
- Javaheri H, Fortes Rey V, Lukowicz P, Stavrou GA, Karolus J, Ghamarnejad O. Assessing the feasibility of using Apple Vision Pro while performing medical precision tasks: controlled user study. JMIR XR Spatial Comput. 2025;2:e73574. [CrossRef]
- Shen K, Li W, Xie Z. Preliminary application of Apple Vision Pro (AVP) in laparoscopic gastrointestinal surgery. Int J Surg. Nov 1, 2025;111(11):7898-7905. [CrossRef] [Medline]
Abbreviations
| AVP: Apple Vision Pro |
| CT: computed tomography |
| HDMI: High-Definition Multimedia Interface |
| MR: mixed-reality |
| NDI: Network Device Interface |
| SSQ: Simulator Sickness Questionnaire |
Edited by Ivan Steenstra; submitted 15.Jun.2026; peer-reviewed by Inigo Tamayo; final revised version received 08.Sep.2026; accepted 09.Sep.2026; published 30.Oct.2026.
Copyright© Shotaro Okachi, Tomoya Horiguchi, Yasuhiro Goto, Michitaka Fujiwara, Kazuyoshi Imaizumi, Naozumi Hashimoto. Originally published in JMIR XR and Spatial Computing (https://xr.jmir.org), 30.Oct.2026.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR XR and Spatial Computing, is properly cited. The complete bibliographic information, a link to the original publication on https://xr.jmir.org/, as well as this copyright and license information must be included.